Details
Download Brochure
LAMSYSTEMS Vertical Laminar flow cabinet BAVnp-01-Laminar-S, Made in Germany
Standard Configuration
Work Chamber
- hinged front sash made of tempered glass; opening, closing and upholding is carried out by gas springs;
- damper against front window closing impact;
- side windows made of tempered glass;
- slanted front surface of the cabinet;
- LED lighting of the work chamber;
- one set of electric sockets in the work chamber (2 pcs. in the right part of the rear wall);
- airflow laminarization screen made of polymer micromesh;
- tabletop made of stainless steel (AISI 304);
- work opening screen for operator protection during UV light operation.
UV light
- UV light is located is the upper part of the work chamber rear wall and is covered with a protective grille made of stainless steel;
Supply and Exhaust Air Filtration System
- supply air coming into the work chamber is cleaned by preliminary filter G4 and supply HEPA filter H14;
- two fans ensuring air supply into the cabinet.
Control System Components
- microprocessor control system;
- pushbutton control panel with LCD screen;
- airflow sensors;
- individual fuse for power supply to the main functions of the cabinet (fan, lighting, UV lamp);
- individual fuse for electric sockets of the work chamber;
- fixed power cord.
Testing Components
- built-in outlet for sampling the air from the high-pressure plenum under the G4 filter for HEPA filter integrity testing.
Stand
- frame stand with footrest;
- caster wheels for cabinet relocation;
- screw supports for cabinet fixation at operating site.
MAIN CHARACTERISTICS
|
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
-for particles of 0.5µm and more
-for particles of 5.0µm and more
|
ISO 5
ISO M (20; ≥5µm); LSAPC
|
|
Class of the installed in the cabinet НЕРА-filters according to EN 1822-1
|
H14
|
|
Prefilter according to EN 779
|
G4
|
|
Average downflow velocity in the work chamber:
- preset velocity, mps
- recommended velocity range for customized setting that guarantees uniformity (laminarity) of the airflow, mps
|
0.4
0.25-0.5
|
|
Illuminance level in the working zone, lux, not less than
|
1000
|
|
Noise level at 1m distance from the cabinet (at centerline of the work opening with the front window sash at working position), dB, not more than
|
55*
|